◘ Preset with 5 programs for wrapped item, unwrapped instrument, liquid, B&D test and free
setting。
◘ Electric swing clutch door is used for easy operation. Manual operation is available in case of
electric failure。
◘ Silicone rubber gasket is used for airtight sealing。
◘ Open instrument can be flash sterilized in 3 minutes while wrapped items can be sterilized in
10 minutes at 134℃。
◘ LCD touch control screen: 5-1/2” to display date, temperature, pressure, cycle stages and
course of time against pressure and temperature。
◘ Automatic pressure control system for pressure and vacuum setting which accuracy is
between 0.1psi。
◘ Printer system is used for the record of sterilizing time, sterilizing No., cycle No., operator
No., temperature set, time set, pressure/vacuum set, maximum & lowest temperature,
actual heating, sterilizing and drying time, total time used and time of open door etc...
In case of out of printing paper, the cycle proceeding will keep on. Heat-sensitive printing
paper is used to print out 19bps and there is no ink need。
◘ HEPA air filter cartridge is used for 0.2μm air purge, flowage more than 95LPM。
Safety devices:
A.Cycle will not start in case door is not locked properly。
B.Door can not be opened in case chamber is under pressure。
C.Door mounted with anti-clamp system, door will stop sliding in case foreign object is
interrupted。
D.Jacket installed with a safety valve to relief overloaded pressure。
E.LCD control screen displays the malfunction such as low water, steam insufficient, air
insufficient and etc。
F.Piping self-detect system alarms at pressure lose more than 1mmHg per minute。
G.Cycles hold in case of power failure。
H.Parameter of programs can be locked by security code。
Specification:
Items
|
SR-SERIES STEAM STERILIZERS |
Model No |
SR-680 |
SR-850 |
Chamber dimensions |
mm |
W610×D1220×H915 |
W610×D1525×H915 |
Overall dimensions |
mm |
W1280×D1575×H1890 |
W1280×D1880×H1890 |
Inner capacity |
L |
680 |
850 |
Design pressure |
kg/cm² |
3.1 |
Chamber material |
SUS |
Stainless steel type 316 |
Jacket material |
SUS |
Stainless steel type 304 |
Door operation |
AUTO |
Swing with clutch |
Setting range time |
min |
0~99 |
Setting range temperature |
℃ |
110~135 |
Chamber dial gauge |
kg/cm² |
76cmHg~6 |
Jacket dial gauge |
kg/cm² |
0~6 |
Controller |
Panel |
Remote control available |
System |
PLC |
Display |
5½” Touch Screen |
Version |
Chinese/English |
Output interface |
RS-232 |
Security |
Door lock |
Above 2PSI auto lock |
Anti-clamp |
Mechanical |
Safety valve |
2.8kg/cm² |
Pressure reducer |
4kg/cm² |
Steam sieve |
Stainless steel strainer |
Auto-detect |
Power-on |
Leak-test |
﹥1mmHg |
Proceeding recall |
Automatic diagnose |
Errors |
Alarm & display |
Codes |
3 sets |
Standards |
CNS/ASME/JIS |
Vacuum |
Drawing |
Water jet / Vacuum pump |
Printer |
Paper |
Heat sensitive type with auto rolling back |
Sources |
Power |
AC220V/380V,3ψ,50/60HZ |
Water |
3~5kg/cm²,1”PT |
Steam |
4~5kg/cm²,1”PT |
Air |
4~5kg/cm² |
Drain |
2½"PT |
Accessories |
Trolley |
1 set of sterilizing carriage and trolley |
Printer paper |
5 rolls |
Manual |
1 copy of operation and maintenance manual |
*Custom-made capacity is available.
*Specifications should be subject to change without prior notice!