◘ Preset with 5 programs for wrapped item, unwrapped instrument, liquid, B&D test and free
setting。
◘ Electric vertical sliding door is used for easy operation. Manual operation is available in case of
electric failure。
◘ Silicone rubber gasket is used for airtight sealing。
◘ Open instrument can be flash sterilized in 3 minutes while wrapped items can be sterilized in 10
minutes at 134℃。
◘ LCD touch control screen: 5-1/2” to display date, temperature, pressure, cycle stages and
course of time against pressure and temperature。
◘ Automatic pressure control system for pressure and vacuum setting which accuracy is between
0.1psi。
◘ Printer system is used for the record of sterilizing time, sterilizing No., cycle No., operator No.,
temperature set, time set, pressure/vacuum set, maximum & lowest temperature, actual
heating, sterilizing and drying time, total time used and time of open door etc... In case of
out of printing paper, the cycle proceeding will keep on。Heat-sensitive printing paper is used
to print out 19bps and there is no ink need。
◘ HEPA air filter cartridge is used for 0.2μm air purge, flowage more than 95LPM。
Safety devices:
A.Cycle will not start in case door is not locked properly。
B.Door can not be opened in case chamber is under pressure。
C.Door mounted with anti-clamp system, door will stop sliding in case foreign object is
interrupted。
D.Jacket installed with a safety valve to relief overloaded pressure。
E. LCD control screen displays the malfunction such as low water, steam insufficient, air
insufficient and etc。
F.Piping self-detect system alarms at pressure lose more than 1mmHg per minute。
G.Cycles hold in case of power failure。
H.Parameter of programs can be locked by security code。
Specification:
Items |
SV-SERIES STEAM STERILIZERS |
Model No. |
SV-110 |
SV-250 |
SV-453 |
Chamber dimensions |
mm |
W410×D660×H410 |
W510×D965×H510 |
W610×D1220×H610 |
Overall dimensions |
mm |
W905×D995×H1690 |
W1005×D1300×H1750 |
W1105×D1555×H1900 |
Inner capacity |
L |
110L |
250L |
453L |
Design pressure |
kg/cm² |
3.1 kg/cm² (0.3Mpa) |
Chamber material |
SUS |
Stainless steel type 316 |
Jacket material |
SUS |
Stainless steel type 304 |
Door operation |
AUTO |
Electric vertical sliding |
Setting range time |
min |
0~99 |
Setting range temperature |
℃ |
110~135 |
Chamber dial gauge |
kg/cm² |
76cmHg~6 |
Jacket dial gauge |
kg/cm² |
0~6 |
Controller |
Panel |
Remote control available |
System |
PLC |
Display |
5½” Touch Screen |
Version |
Chinese/English |
Output interface |
RS-232 |
Security
|
Door lock |
Above 2PSI auto lock |
Anti-clamp |
Mechanical |
Safety valve |
2.8kg/cm² |
Pressure Reducer |
4kg/cm² |
Steam sieve |
Stainless steel strainer |
Auto-detect |
Power-on |
Leak teat |
﹥1mmHg |
Proceeding recall |
Automatic diagnose |
Errors |
Alarm and display |
Codes |
3set |
Standard |
CNS/ASME/JIS |
Vacuum |
Drawing |
Water jet / Vacuum pump |
Printer |
Paper |
Heat sensitive type with auto rolling back |
Sources |
Power |
AC220V/380V,3ψ,50/60HZ |
Water |
3~5kg/cm²,1”PT |
Steam |
4~5kg/cm²,1”PT |
Air |
4~5kg/cm² |
Drain |
2"PT |
Accessories |
Loading appliances |
1 removable rack w/ 2 baskets |
1 carriage w/ 1 loading cart |
1 carriage w/ 1 loading cart |
Printing paper |
5 rolls |
Manual |
1pc |
*Custom-made capacity is available.